Model: BM-200C

Fill in the form to get more product information

Product Introduction

Product Application

CPAP system is used for newborn baby with stronger spontaneous breath and slightly respiratory failure, and applies to pediatric clinics for NCPAP oxygen uptake.

Product Performance and features

1. CPAP ventilation mode keeps children spontaneous breath and reduces children breathing work, continues to provide CPAP mode of constant flow and limited pressure for newborn baby.

2. Delivery air/O2 mixture with properly adjusted O2 concentration and continuous positive pressure to premature baby, newborn baby and child, which will highly improve the safety and efficacy of oxygen therapy.

3. Equipped with built-in oxygen/air blender, ensuring O2 concentration adjustment to be stable.

4. High quality oxygen sensor can monitor real-time O2 concentration, alarm over limitation, ensure safety of breathing O2.

5. Reliable CPAP control valve and advanced pressure monitoring system improve CPAP adjustment precision.

6. 3″ display.

Product Working condition

Air source: O2, AIR

Air source pressure: 280 kPa ~ 600 kPa

Power supply: ~220 V±22 V, 50 HZ with air compressor, 100-240V, 50/60HZ (without air compressor)

Input power: 900VA (with air compressor), 250VA (without air compressor)

Ventilation Mode: CPAP

Main technical parameters

Adjusting Parameter

Inhaled oxygen concentration: 21%~100%

CPAP: 0cmH2O~15 cmH2O

Continuous flow: 2L/min~20L/min

Monitoring Parameter

Inhaled oxygen concentration: 15% ~ 10 %

CPAP: 0 cmH2O ~ 60 cmH2O

Continuous flow: 2 L/min ~ 20 L/min

Alarm and protection

The AC power cut-off alarm

Low air (oxygen) input pressure alarm

Air / oxygen input unbalanced pressure alarm

High oxygen concentration alarm

low oxygen concentration alarm

Pressure limit